1. STATEMENT OF WORK The contractor shall furnish all necessary services, labor, travel, transportation, materials, shipping, parts, connections, and equipment. This includes the provision and on-site installation of at least 5 AMSCO 50 Series Reprocessing Sinks at 4801 Linwood Blvd, Kansas City MO 64128. These reprocessing sinks are required by the Sterile Processing Services (SPS) to facilitate the manual cleaning of sterilization-compatible reusable medical devices (RMD) in invasive procedures. Their use will support a range of medical procedures and surgeries, including those in the operating room, dental services, procedural gastroenterology, procedural pulmonology and surgical clinics. The contractor bears the responsibility for the removal and disposal of all debris and packaging materials resulting from the reprocessing sinks installation. Prior to equipment utilization, the contractor is required to provide training to clinical staff, with all duty tours for the sterile processing department being documented. The scope of this contract comprises five AMSCO 50 Series Reprocessing Sinks. Associated services include complete installation components, execution of a post-installation check-out test, and implementation of any necessary adjustments. 2. SPECIFICATIONS - Minimum of five 3 Bay 97.5 x 28 x 33 - 41 Adjustable Height Reprocessing Sinks. - Installer Must Maintain a Minimum Gap of .875 To All Adjacent Walls as a Pinch Point Protection. - Unit Must Be Installed Against a Wall, For Back-to-Back Installation a Pony Wall is Required. - 304 Stainless Steel Construction.
- Utility requirements (plumbing) - Hot Water (HW)
- Operating Flow Rate- 1.6GPM @ MAX PSI
- Water Quality- Steris recommends the following water quality parameters: - 100-200 PPM (500PPM MAX) TOTAL DISOLVED SOLIDS. - 71-120 PPM (180PPM MAX) TOTAL ALKALINITY. - 6.8-7.0PH (7.5 MAX). - 0.1-0.5PPM (1.0) TOTAL SILICA. - Cold Water (CW)
- Operating Flow Rate- 1.6GPM @ MAX PSI
- Water Quality- Steris Recommends The Following Water Quality Parameters: - 100-200 PPM (500PPM MAX) TOTAL DISOLVED SOLIDS. - 71-120 PPM (180PPM MAX) TOTAL ALKALINITY. - 6.8-7.0PH (7.5 MAX). - 0.1-0.5PPM (1.0) TOTAL SILICA - PURE WATER (PW)
- Operating Flow Rate- .25 TO 2.0 GPM 15%
- WATER GUN (WG) - Connection type/size- 3/8 Compression
- Water Quality- Recommendations listed under #5 of (HW & CW) - Air Gun (AG)
- Tip 1-5- Tip consumption at 30 PSI (Tip 4 and 5 must be operated at 30 PSI or less to comply with OSHA Directive. - Basin Drain (BD) - Connection Type- 1-1/2 ABS P-TRAP (INCLUDED), 2 Copper P-Trap (Accessory) - Drain Rate At 20 GPM Each Basin (Estimated)
- When Traps Are Not Required, Elbows Can Be Purchased Locally. - Electricity- 120V, 15A GFCI Protected Outlet
- Weight Capacity- 562LBS (2500 N) Each - Adjustment Range- 8.0 (203.2) - Lift Speed- Up To 3/8 (10) Per Second - Voltage- 120VAC, 3A With Overload Protection
- Operating Dimensions L x W x H- 97.5 x 28 x 44-9/16 TO 52-9/16 [2477 x 711 x 1132 TO 1412] - Concrete Floor Must Remain Free of Defects in These Areas for Installation of Seismic Anchors (Not Provided by Steris). - Refer to Equipment Specific Seismic Drawing for More Information. - The Reprocessing Sinks must adhere to industry standards for AAMI/ ANSI, OSHA, etc. - Documented FDA approval, as applicable. 3. REQUIREMENTS - The Contractor is responsible for conducting site visits to verify existing conditions, material quantities, installation instructions, and to identify any discrepancies or conflicting requirements within the Statements of Work before preparing the Final Proposal. - The Contractor shall provide and install five AMSCO 50 Series Reprocessing Sinks to the VA Medical Center located at 4801 Linwood Kansas City, Mo 64128. - The Contractor must ensure that all repairs, preventative maintenance and remedial maintenance provided during this contract's timeframe meet or exceed the OEM'S specifications. - The contractor shall employ ONLY FULLY QUALIFIED TECHNICIANS/ENGINEERS for all work performed under this contract. - The contractor shall ensure that only FULLY QUALIFIED TECHNICIANS/ENGINEERS are employed under this contract. - The Contractor shall maintain the highest standard of professional capability and workmanship throughout the life of this contract. 4. EQUIPMENT MAINTENANCE & RESPONSE TIMES The vendor shall provide full-service maintenance through a service agreement for minimum 1 year post installation. The vendor shall provide all additional recurring and preventive maintenance activities: - Emergency response for equipment and MSPS failures. The vendor shall respond to the facility for any equipment failure resulting in a total or partial loss of the equipment installed within or any portion of the mobile sterile processing system within 4 hours from the time of VAMC contact. If a permanent solution cannot be resolved within 24 hours of initial failure, a temporary solution shall be operational within 24 hours of initial failure. An agreed permanent solution shall be planned, and replacements ordered within 5 business days of initial failure. 5. SAFETY The contractor/vendor shall always, ensure that all work provided complies with all local, State, and Federal rules and regulations pertaining to workplace safety. The contractor shall take all proper safety and health precautions to protect the work, the workers, the public and the property (OSHA). Installs and construction work will be in accordance with all applicable codes to include the most current codes National Fire Protection Association (NFPA), and the VA Design Guide Manuals that are applicable to the work. If the design guide is equivalent or more stringent then the mentioned codes, the VA Design Guide takes precedence. 6. DOCUMENTATION The contractor shall furnish to the KCVA Medical Center Biomedical Engineering department legible hard copies and electronic copies of service reports of the maintenance indicating details of the maintenance performed upon completion of all work. 7. PLACE OF PERFORMANCE Kansas City Veteran Administration Medical Center, 4801 Linwood Blvd. Kansas City 64128 8. PERFORMANCE MONITORING The contractor's performance will be monitored by the Contracting Officer (CO), Contracting Officer Representative (COR), Infection Control Risk Assessment Office, and Safety Office during the rental agreement period. Coordinate all administrative and construction requirements exclusively with or through assigned Contracting Officer Representative (COR). The Construction Safety Officer, Construction Safety Committee, and COR will review the operations and site construction conditions of the project to confirm that all Safety regulations relating to construction and healthcare facilities are being adhered to. Daily and weekly safety rounds will be conducted. For non-compliance safety related issues, a safety report will be sent to the Contractor to correct within 14 days. Non-compliance may result in job-site suspense until compliance is met at the contractor's expense. 9. HOURS OF OPERATION The contractor will perform services required during the hours of 7:30 am 4:00 pm, Monday through Friday or as directed by the designated COR.
Background The Department of Veterans Affairs is issuing this contract to provide and install AMSCO 50 Series Reprocessing Sinks at the Kansas City Veterans Administration Medical Center. The goal of this contract is to facilitate the manual cleaning of sterilization-compatible reusable medical devices in various medical procedures, thereby supporting the mission of the Sterile Processing Services (SPS). Work Details The contractor shall furnish all necessary services, labor, travel, transportation, materials, shipping, parts, connections, and equipment for the installation of at least 5 AMSCO 50 Series Reprocessing Sinks. The specifications for the sinks include: - Minimum of five 3 Bay 97.5” x 28” x 33”-41” Adjustable Height Reprocessing Sinks. - Construction must be 304 Stainless Steel. - Installation requirements include maintaining a minimum gap of .875” to all adjacent walls and ensuring that the wall is free of obstructions to 54”. - Utility requirements include hot water with specific plumbing connections and pressure ranges. - The contractor is responsible for training clinical staff prior to equipment utilization and must ensure compliance with industry standards for AAMI/ANSI, OSHA, etc. Period of Performance The period of performance will commence upon contract award and will include a service agreement for a minimum of one year post-installation. Place of Performance Kansas City Veteran Administration Medical Center, 4801 Linwood Blvd, Kansas City MO 64128. Bidder Requirements This solicitation is set aside for Small Business participation. Bidders must be authorized distributors or manufacturers of the AMSCO products and are required to submit an Authorized Distributor Letter from the Original Equipment Manufacturer (OEM) along with their quote. Incumbent Analysis (see Incumbents section for more detail) This opportunity appears to have a confirmed incumbent and/or closely related prior contract based on V797D60613-36C24225F0192 awarded to Steris Corporation for $49,991 (08/25/25–11/24/25). That prior work involved purchase and installation of STERIS AMSCO reprocessing sinks, including deinstallation of existing AMSCO 50 sinks, which aligns with the current requirement to provide and install additional AMSCO 50 Series reprocessing sinks at the same Kansas City VA Medical Center. The focus on AMSCO reprocessing sink installation/deinstallation suggests the current effort may be a continuation or expansion of the facility’s sterile processing sink infrastructure.
| Contact name | Jon Lutmer |
|---|---|
| Contact email | jonathan.lutmer@va.gov |
| Contact phone | (410) 446-3862 |
Agency: VISN 15: Heartland Network
Location: Missouri
NAICS: 339112
Agency: Department of the Army
Location: Missouri
NAICS: 236220
Agency: VISN 15: Heartland Network
Location: Missouri
NAICS: 339113
Agency: VISN 15: Heartland Network
Location: Missouri
NAICS: 339113