| Set Aside | SDVOSBS |
| NAICS | 339113 - Surgical Appliance and Supplies Manufacturing |
| PSC | 1190 - Specialized Test And Handling Equipment, Nuclear Ordnance |
| Location | Colorado Springs, CO 80913 United States |
System must allow clinicians to adjust the focal point and working distance directlyfrom the center sterile handle without moving the light head or leaving the sterilefield. The light head must utilize vertically segmented reflectors to capture upward-directed light and redirect it downward, creating a crisp, homogenous spot thatminimizes shadows. The system must offer dual selectable color temperatures (specifically 4100K softwhite and 4500K bright white) adjustable for specific procedure types.
The lights must deliver a maximum intensity of 160,000 lux and offer light headdiameter options of 24 inches or 30 inches to fill the surgical cavity with seamlesslight. The lighting architecture must integrate a manual focus mode of high-intensity LEDscombined with a centrally positioned ring of fixed-focus LEDs. The lights must be explicitly compatible with and ready for integration with flat-screen displays, booms, and HD surgical cameras.
The suspension and arm system must provide at least 110 degrees of vertical travelto allow steep downward positioning. All axes of the lighting fixture must feature 360-degree continuous rotation withouthard spots. The optical design must allow the fixtures to be placed further away from the patientwhile maintaining precise detail.
Background The contract is issued by the Department of Defense, specifically for the procurement of L&D Surgical Lights. The mission of this procurement is to enhance surgical procedures by providing advanced lighting solutions that meet clinical and technological requirements, ensuring optimal performance in sterile environments. Work Details The contract requires the delivery of L&D Surgical Lights with the following specifications: - Clinical Requirements: - System must allow clinicians to adjust the focal point and working distance directly from the center sterile handle without moving the light head or leaving the sterile field. - The light head must utilize vertically segmented reflectors to minimize shadows and create a crisp, homogenous spot. - Must offer dual selectable color temperatures (4100K soft white and 4500K bright white). - Technological Requirements: - Maximum intensity of 160,000 lux with light head diameter options of 24 inches or 30 inches. - Integration with flat-screen displays, booms, and HD surgical cameras is required. - System Requirements: - Suspension and arm system must provide at least 110 degrees of vertical travel. - All axes must feature 360-degree continuous rotation without hard spots. - Optical design must allow placement further from the patient while maintaining detail. Period of Performance The geographic location(s) the product will be delivered to or performed at Bidder Requirements Bidders must be authorized distributors or manufacturers of the specified surgical lights. The solicitation is set aside for Service-Disabled Veteran-Owned Small Businesses (SDVOSB Sole Source).
| Contact name | Joseph Flores |
|---|---|
| Contact email | joseph.a.flores.civ@health.mil |
| Contact phone | None |
Agency: Arapahoe County [CO]
Location: Colorado
NAICS: Not Provided
Agency: City of Brighton [CO]
Location: Colorado
NAICS: Not Provided
Agency: Colorado Department of Corrections Administration (DOC) [CO]
Location: Colorado
NAICS: Not Provided
Agency: Denver Public Library [CO]
Location: Colorado
NAICS: Not Provided